The uncorrected FEI TitanX 60-300 microscope at NCEM is equipped to do fast energy dispersive X-ray mapping (EDS), 4D-STEM, S/ TEM imaging and tomography. All TEM holders that are compatible with FEI microscopes can be used for heating, cooling and other in-situ experiments. The ability to switch between 60, 80, 120, 200, and 300kV during one TEM session gives the user flexibility to optimize experimental conditions – balancing sample damage, spatial resolution, and signal-to-noise.
Images are acquired with a Gatan UltraScan 1000 (2k x 2k) Camera for high-resolution TEM, a Gatan Orius 830 (2k x 2k) for diffraction and movies (30 frames per second), or a ADF detector for STEM. The Bruker windowless EDS detector with a solid angle of 0.7 steradians enables high count rates with minimal dead time. The ~5 mm pole gap of the FEI Super-twin objective lens is compatible with a variety of holders. Advanced experimental techniques such as 4D-STEM and EDS with tomography are being implemented on the TitanX.
Most TEM holders that are compatible with FEI microscopes are available. See the TEM Holder page for more details.
- FEI low background double-tilt (DT) for EDS
- Gatan 698 Elsa Cryo-Transfer Tomography, EDS
- Gatan 915 DT Cryo-Transfer, EDS
- Gatan 636 Cryo
- Gatan 671 Cryo Straining
- Gatan 654 Straining
- Gatan 652-TA DT Heating to 1000°C
- Hummingbird Tomography, also EDS
- Hysitron PI-95 for mechanical compression, tension, biasing
- Gatan 648 DT Air-Free, also EDS and tomography
- Fischione Single tilt Air-Free , also EDS
- Hummingbird single tilt biasing holder for user-defined MEMs devices
- Protochips Aduro/Fusion Heating/Biasing DT
- Protochips Aduro Heating/Biasing Single tilt
- FEI single tilt for general use
- FEI low background double tilt with hexring for magnetic samples
General Specifications
Accelerating Voltage | 60, 80, 120, 200, and 300 kV |
Spherical Aberration Cs | 1.2 mm |
Chromatic Aberration Cc: | 1.2 mm |
Detectors
EDS | Four windowless SDD indicated solid angle of 0.7 steradians and 140 eV energy resolution. |
STEM | Gatan ADF |
HRTEM | Gatan UltraScan 1000 2k x 2k CCD; 2 fps |
Diffraction/Movies | Gatan Orius 830 2k x 2k CCD; 30 fps |
Specimen Stages
FEI low background double tilt for EDS | – | ±40°/±40° |
Hummingbird Tomography holder for tomography and EDS | – | ±70°/±70° |
Protochips double tilt holder for in-situ heating and biasing and EDS | 23°C-1000°C | ±40°/±40° |
Protochips single tilt holder for in-situ heating and biasing | 23°C-1000°C | ±30°/±30° |
FEI low background double tilt for general use | – | ±30°/±30° |
FEI single tilt for general use | – | ± 30° |
Gatan 652-Ta double tilt heating holder | 23°C-1000°C | ±30°/±30° |
Gatan 636-DH low background liquid nitrogen cooling holder | -170°C/ 23°C | ±30/±30° |